Eagle Home Products Eagle Series User Manual Page 59

  • Download
  • Add to my manuals
  • Print
  • Page
    / 80
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 58
Eagle Instruction Manual Appendix B: Methane Elimination 53
2. Allow 2 minutes for the combustibles sensor to stabilize.
3. Perform the demand zero procedure as described in “Preparing
for Calibration” on page 41.
Monitoring combustible gases other than hexane
Use Table 7 to determine the concentration of combustible gases other
than hexane. This table is based on Eagle’s in methane elimination
mode (methane elimination switch set to HEX ON) and calibrated to
hexane. Multiply the display reading by the factor in the appropriate
column. For example, if you are using the Eagle to detect toluene and
the display reads 10% LEL, the actual toluene reading is 10% x 0.90 =
9% LEL toluene.
To determine the concentration of other combustible gases with the
Eagle in full response mode, see Table 4 on page 12.
WARNING: The Eagle’s alarms are initiated by the DISPLAY reading
not the FACTORED reading. If you are monitoring for
toluene as in the above example and the low alarm is set
for 10% LEL, the Eagle will initiate a low alarm at 9%
LEL toluene (display reading of 10% LEL).
Table 7: Methane Elimination Mode Conversion Factors (Hexane Calibration)
Target Gas
LEL Factor
(Hexane Calibration)
PPM Factor
(Hexane Calibration)
Benzene 0.90 0.98
Ethane Low response Low response
Ethanol 0.75 2.25
Ethylene 0.53 0.57
Hexane 1.00 1.00
Hydrogen 0.55 2.00
IPA 0.98 1.77
Isobutane 0.74 1.21
MEK 0.91 1.49
Methane No response No response
Methanol 0.91 4.56
Propane Low response Low response
Propylene 0.69 1.26
Toluene 0.90 0.98
Page view 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 79 80

Comments to this Manuals

No comments