Eagle Home Products Eagle Series User Manual Page 10

  • Download
  • Add to my manuals
  • Print
  • Page
    / 80
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 9
4 Specifications Eagle Instruction Manual
Table 2 lists specifications for the Eagle’s standard sensors. Your Eagle
model may not include all of the sensors listed below. The alarm
settings are user-adjustable (see “Updating the Alarm Point Settings”
on page 32.)
Table 2: Standard Sensor Specifications
Combustible
Gas (%LEL
1
)
Combustible
Gas (PPM
2
)
Oxygen
Hydrogen
Sulfide
Carbon
Monoxide
Range 0 to 100% LEL Depends on
target gas
4
0 to 40% O
2
0 to 100
ppm
0 to 500
ppm
Alarm 1 10% LEL 5000 ppm 19.5% O
2
(decreasing)
10.0 ppm 25 ppm
Alarm 2 50% LEL 25,000 ppm 23.5% O
2
(increasing)
30.0 ppm 50 ppm
TWA Alarm N/A N/A N/A 10.0 ppm 25 ppm
STEL Alarm N/A N/A N/A 15.0 ppm 400 ppm
Detection
Principle
Catalytic
combustion
Catalytic
combustion
Electro-
chemical
Electro-
chemical
Electro-
chemical
Response Time
(to 90%)
5
30 seconds 30 seconds 30 seconds 30 seconds 30 seconds
Accuracy
(of fullscale)
± 5% ± 5% ± 5% ± 5% ± 5%
1 LEL (Lower Explosive Limit)
2 PPM (Parts Per Million)
3 Alarms settings are user adjustable. See “Updating the Alarm Point Settings” on page 32.
4 The PPM range represents the same range as 0 to 100% LEL for that gas. For example, 100% LEL for methane
= 5% by volume = 50,000 PPM. Therefore, the PPM range for methane is 0 to 50,000.
5 With the Eagle’s standard hose and probe attached.
Page view 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 79 80

Comments to this Manuals

No comments